ED Decision 2016/029/R
The applicable standard is SAE AS8025A, Passenger Oxygen Mask, dated (revised) January 1999 and shall be modified as follows:
SAE AS8025A Paragraph |
Action |
Section 1, SCOPE |
To be disregarded. |
Paragraph 3.2, Deviations |
To be disregarded. |
Paragraph 3.3.1, General |
Shall be revised: ‘Construct the device, including packaging, of materials that will not contribute significantly to fire propagation and that comply with CS 25.853(a). Mask materials typically used should meet CS-25 Appendix F, Part I(a)(1)(ii) and/or Part I(a)(1)(iv).’ |
Paragraph 3.3.3, Cleaning and sterilization |
Shall be revised: ‘Cleaning and Sterilizing: The material of the oxygen mask shall permit cleaning and sterilizing without adverse effects, and without major disassembly. The cleaning method must be either manufacturer-recommended, or according to SAE ARP 1176, Oxygen System Component Cleaning and Packaging. Cleaning and sterilizing procedures shall be included in the CMM.’ |
Paragraph 3.3.4, Elastomer Components |
The following sentence shall be added: ‘Life limits and inspection procedures shall be included in the CMM.’ |
Paragraph 3.11, Identification Markings |
To be Disregarded. Marking requirements are specified in paragraph 4 of this ETSO. |
Paragraph 4.5.2 |
Flow indication must comply with AS 916B, Oxygen Flow Indicators, as applicable |
[Amdt ETSO/12]
ETSO-C64b specifies SAE AS8025A with modifications for passenger oxygen mask assemblies, including fire propagation compliance with CS 25.853(a), cleaning per SAE ARP 1176, and CMM inclusion of life limits and inspection